2 GENERAL SPECIFICATIONS

2.1 Measuring elements

: pH glass electrode

Sodium glass electrode
Solid Platinum electrode
Pt1000 temperature sensor

2.2 Wetted parts

Measuring sensor

: Glass

Earth pin

: Solid Platinum

2.3 Functional specifications (at 25°C)

Isothermal point

: pH 7, pNa 0

Reference system

: pNa, salt sensitive reference glass

Glass impedance - pH

: 750 MΩ nominal

- pNa : 750 MΩ nominal

Temperature element

: Pt1000 to IEC 751

Asymmetry potential (zero) : 0 ± 15 mV

Slope

: > 90 % (of theoretical value)

Note: The SC24V temperature sensor is designed for cell compensation and for indication.

It is

NOT designed for process temperature control.

2.4 Dynamic specifications

Response time pH

: t

90

< 15 sec. (for 7 to 4 pH step)

Response time temperature

: t

90

< 1.5 min. (for 10 °C step)

Stabilization time pH

: < 2 min. (for 0.02 pH unit during 10 sec.)

2.5 Operating range

pH

: 2 to 14 (refer to Table 3)

Temperature

: 10 ºC to 120 °C (50 ºF to 248 °F)

Pressure

: 0 to 10 bar (0 to 142 PSIG)

Conductivity

: > 50 µS/cm

Table 3: pH operating range as function of the ionic strength (pH≥pNa+2)

pNa

Ionic strength [mol ion+ per liter]

pH operating range

0

1

2 to 14

1

0.1

3 to 14

2

0.01

4 to 14

3

0.001

5 to 14

4

0.0001

6 to 14

Note: The pH operating range is given at room temperature, but at high process temperatures the

lifetime of the sensor will be seriously shortened. For pH values > 12 the accuracy of the pH
measurement will change significantly by the alkali error.