Yokogawa PH18 Differential pH Sensor User Manual
Page 12

12
IM 12B6J4-E-A
Example A. Straight weld-in adapter through a vessel
wall.
Example B. 1” Screw-in adapter with existing pipe
nipple through a vessel wall.
Example C. Angled weld-in adapter in large bore pipe.
Example D. Angled weld-in adapter mounted in a
bend
Example E. Screw or weld-in adapter mounted in a
large bore pipe.
Example F. Screw or weld-in adapter mounted on top
Fig. 11 The primary adapter is used to make a
connection to the process used by the transmitter as
the liquid earth (solution ground). This is needed for
optimum stability of the measurement and is used in
the impedance checking circuit.
Note 1:
When measuring in plastic tanks or pipes, ensure that the adapter is
wetted by the process. Avoid installations where an air pocket can
be created (see fig 12 F). This isolates the adapter, and hence loses
the liquid earth connection.
Note 2:
Flow rate from the side of the sensor (Fig 11) should not exceed 2
meters/second in low viscosity fluids. In high viscosity fluids (> 5cP)
use only installation as
shown in Fig. 12 (D).
Fig. 12. Model 18 elements
IM 12B6J4-E-A
9
3-5. PH18 Installation examples
Note 1:
When measuring in plastic tanks
or pipes, ensure that the adapter is
wetted by the process. Avoid
installations where an air pocket can be
created (see fig 12 F). This isolates the
adapter, and hence loses the liquid earth
connection.
Note 2:
Flow rate from the side of the sensor (Fig
11) should not exceed 2 meters/second
in low viscosity fluids. In high viscosity
fluids (> 5cP) use only installation as
shown in Fig. 12 (D).
Example A. Straight weld-in adapter
through a vessel wall.
Example B. 1” Screw-in adapter with
existing pipe nipple through a
vessel wall.
Example C. Angled weld-in adapter in
large bore pipe.
Example D. Angled weld-in adapter
mounted in a bend
Example E. Screw or weld-in adapter
mounted in a large bore pipe.
Example F. Screw or weld-in adapter
mounted on top of a large
bore pipe
Fig. 11 The primary adapter is used to
make a connection to the process used by
the transmitter as the liquid earth (solution
ground). This is needed for optimum
stability of the measurement and is used in
the impedance checking circuit.
Socket for connection
cable
Primary adapter
pH enamel
Reference enamel
Fig. 12. Model 18 elements
3-5. PH18 Installation examples