5.1 Connections
The FU24 sensor is provided with a fixed cable or Variopin (VP) connector. The connections
of both the wire as the VP connector are given in Figure 4.

Figure 4. Cable marking and pin locations

5.2 Definition EXA/FLXA terminal numbers to FU24 sensor

FU24-05/10 wire #

pH

ORP

EXA/FLXA

terminal

#

pH+ORP

ORP

pH compensated

pH+rH

11 11 11 11

12 12 12 12

13 13 13 15

14

14

15 15 14 14

16

16

17

16

FU24-VP with WU10-V wire color

pH

ORP

EXA/FLXA

terminal

#

pH+ORP

ORP

pH compensated

pH+rH

11 Red Red Red

12 Blue Blue Blue

13

Brown

Brown

Clear

14

Yellow

15

Clear

Yellow

Yellow

16

Black

17

Black

Note: For ORP measurement the EXA/FLXA terminals 13 and 14 should be interconnected.

For ORP pH compensated measurement the EXA/FLXA terminals 14 and 15 should
be interconnected.

5. WIRING

(pH) A

(Ref.) B

(pH shield) C

(ORP/LE)

D

E (Temp-1)

F (Temp-2)

VP-connector

(pH) Clear

(Ref.) Brown

(ORP/LE) Yellow

(Temp-1) Red

(Temp-2) Blue

(pH shield) Black

FU24

WU10-V

MODEL CODE

Serie nummer

11/red, (Temp-1)

ø6.5

12/blue, (Temp-2)

13/black, (Ref.)

15/clear, (pH)

16/brown, (pH shield)

14/yellow, (ORP/LE)

MODEL CODE

Serie nummer

ш6.5