2 connections of vp dual coax cable – Yokogawa SC24V Differential pH ORP Electrode User Manual
Page 9

IM 12B6W2-02E-E
9
Red (E)
Brown (B)
Transparant (A)
Yellow (G)
Blue (F)
White (D)
Transparant (C)
Green (H)
Black
Cable
A
B
C D E
F
brown
white
red
blue
brown
Connector
green
yellow
white
H G
3.2 Connections of VP dual coax cable
The numbers in the table refer to the numbers of the terminals in the instrument.
A B C D E F G H
SC24V pH + pNA + ORP(rH)
15 16 13 17 11 12 14 -
(63)
(3)
none
FU24/FU20: pH + ORP(rH)
15 13 16 14 11 12 - -
(63)
(3) -
FU24/FU20: pH / ref
15 13 16 14 11 12 - -
(63)
(3) -
FU24/FU20: ORP / ref
- 13 16 15 11 12 - -
(63)
(3)
13-14
FU24/FU20: ORP / pH
13 - 16 15 11 12 - -
(63)
(3)
15-14
Polagraphic DO sensor (OXYgold)
17 18 - 14 11 12 - -
(63)
(3) -
Galvanic DO sensor
13 15 - (14) 11 12 - -
(63)
(3) -
2-electrode conductivity sensor
15 16 13 14 11 12 - -
(63) 13-14
(3)
15-16
4-electrode conductivity sensor
16 15 13 14 11 12 - -
(63)
(3) -
Table 4
The numbers (11- 17) in the table refer to the numbers on the terminals in EXA Instruments.
Other instruments are indicated in the following tables
Transparent (Brown core)
Brown
Transparent (White core)
White
Red
Blue
Yellow
Green
Black
Jumper
Instrument
PH202
PH402
PH450
PH150
DO202
DO402
SC402
SC202
SC450
SC150
Figure 3
Figure 4