IM 12B6W2-02E-E
8
3. Installation
3.1 Connections of VP single coax cable
The numbers in the table refer to the numbers of the terminals in the instrument.
A B C D E F Jumper Instrument
FU24/FU20: pH+ORP(rH) 15
13 16 14
11
12
FU24/FU20: pH / ref
15
13 16 14
11
12
PH202 / PH402
FU24/FU20: ORP / ref
-
13 16 15
11
12
13-14 PH450 / PH150
FU24/FU20: ORP / pH
13
- 16 15
11
12
15-14
Polagraphic DO sensor
(OXYgold)
17
18
-
14
11
12
-
DO202
Galvanic DO sensor
13
15
- (14) 11
12
-
DO402
2-electrode
conductivity
13-14
sensor
15
16
13
14
11
12
15-16 SC402 / SC202
4-electrode conductivity
SC450 / SC150
sensor
16
15
13
14
11
12
-
Tabel 2
The numbers (11- 17) in the table refer to the numbers on the terminals in EXA Instruments.
Other instruments are indicated in the following tables
PH100
function
pin ID
Color
name
number
pH
A
Transparent GE
12
Ref
B
Brown
RE
14
C
Black
Ground el.
D
Yellow
SE
16
TC
E
Red
T1
17
TC
F
Blue
T2
18
Tabel 3
A
B
D (yellow)
E (red)
F (blue)
C (black)
(transparant)
(brown)
Cable
A
B
C D E
F
brown
yellow
black
red
blue
transparant
Connector
Transparant
Brown
Black
Yellow
Red
Blue
Figure 1
Figure 2