2 entering the batch parameters – Yokogawa 414 Batch Controller User Manual
Page 36

34 Calibration
5.2 ENTERING THE BATCH PARAMETERS
Step
Display
Description
Text
Ref
1
BATCH
OPTION
TEST
END
CAL
Enter Batch Parameters.
Option (if installed)
Check Input Signals.
Exit to normal operation.
Program Setup Parameters.
5.3
5.4
5.1
The following steps are displayed if BATCH is selected.
2
BATCH L
xxxxxx
Maximum Batch Size which can be entered.
Set to 0 if no limit on batch size.
3.2
3
AUTO S
Off
On
xx:xx
Automatic restart feature.
Disable.
Enable.
If enabled, automatically restarts the batch
xx:xx (mins:sec) after the end of the last batch.
3.2.4
4
START. T
xx:xx
Slow start time.
Time, in (minutes:seconds), when Relay 2 will
energise once the batch has started.
3.2
5
PREST
xxxx
Prestop Quantity.
Quantity at which Relay 2 will de-energise
before the end of the batch. (eg. If the batch
quantity is 100 litres and Prest is 2 litres, relay
2 will de-energise after 98 litres.)
3.2