Please use the attached sheets for the pages listed below in IM 01E20D01-01E (10th).
Page and Item
Page 1-2, 1-3
1.1 Using the Magnetic Flowmeter Safety
Page 1-4, 1-5, 1-6
1.3 Combination Remote Converters
1.4 ATEX Documentation
Page 3-3
Figure 3.2.1, Lifting Flowmeter
Page 4-4
(3) Precautions for Wiring of Power
Supply Cables
Page 4-6
Figure 4.1.10, 4 to 20 mA DC Output
Connection
Figure 4.1.12, Active Pulse Output Con-
nection (Optional code EM)
Page 4-8
4.2 Wiring the Remote Flowtube
Page 6-10
Withstand Voltage
Page 6-11
Safety Requirement Standards
Page 8-5
8.3 CSA
Contents of Correction
Added the warning notes in (1) installation, (3)
Operation, and (4) Maintenance.
Changed the page number.
Changed the page number.
Added another horizontal position figure.
Changed the numbers of “IEC” in warning note.
Changed the value of maximum load resistance.
Changed the figure.
Added the important note on “wiring”.
Changed the values of the withstand voltage.
Added the number of “EN” standard.
Changed the number of Applicable Standard
in (1) Technical Data.