User’s manual – Yokogawa EJA115 User Manual
Page 4
User’s
Manual
Sep. 15, 2010
Yokogawa Electric Corporation
EJA Series
Differential Pressure and
Pressure Transmitters
Please use this manual change for the manuals listed below.
1. Items to be Changed and Applicable Manuals
IM No. and Edition
Products
IM 01C21B01-01E (11th)
EJA110A, EJA120A, and EJA130A
(1), (2)
IM 01C21C01-01E (11th)
EJA210A and EJA220A
IM 01C21F01-01E (8th)
EJA510A and EJA530A
IM 01C22H01-01E (17th)
EJA118W, EJA118N and EJA118Y
IM 01C22T02-01E (10th)
EJA Series Fieldbus Communication Type
Item to be applied
(Refer to '2.Contents of Changes')
IM 01C21D01-01E (11th)
EJA310A, EJA430A and EJA440A
IM 01C22J01-01E (17th)
EJA438W and EJA438N
IM 01C22K01-01E (17th)
EJA115
IM 01C22T03-00E (5th)
EJA Series PROFIBUS Communication Type
IM 01C22A01-01E (2nd)
EJA Series Installation Manual
(1)
(1)
(1)
(1), (2)
(1), (2)
(1), (2)
(1), (2)
(1), (2)
(1), (2)
10-007-1.EPS
2. Contents of Changes
(1)
For all the statements for the ambient temperataure limit of the prodcut in the applicable manuals, add the
following condition;
The lower ambient temperature limit shall be -15
°
C (5
°
F) for option code /HE unless other
specifications restrict the llimit.
(2)
Add the following item in the "Optional Specfications" table in the chapter of "General Specifications" in each
manual.
Description
10-007-2.EPS
Item
Code
Fluoro-rubber O-ring
All O-rings of amplifier housing. Lower limit of ambient temperature : –15
°C(5°F)
HE
Manual Change No. 10-007