Appendixes
ACCU-CHEK Spirit Insulin Pump
171
Technical Standards on Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The ACCU-CHEK Spirit insulin pump is intended for use in the electromagnetic
environment specified below. The customer or the user of the ACCU-CHEK
Spirit insulin pump should assure that it is used in such an environment.
IEC 60601
Compliance
Electromagnetic
Immunity Test Test Level
Level
Environment – Guidance
Electrostatic
± 6 kV contact ± 8 kV contact
Avoid any contact with
discharge (ESD) ± 8 kV air
± 15 kV air
synthetic materials. A
IEC 61000-4-2
(IEC 60601-2-24) relative humidity of 10% will
allow ESD to be more likely.
Power frequency 3 A/m 400 A/m 400 A/m
Power frequency magnetic
(50/60Hz)
fields should be at levels
magnetic field
characteristic of a typical
IEC 61000-4-8
(IEC 60601-2-24) location in a typical
commercial or hospital
environment.
Conducted RF
3 Vrms
Not applicable
IEC 61000-4-8
150 kHz to
80 MHz