Automatic Off
Automatic off is a safety feature that stops insulin delivery and triggers an
error E3: AUTOMATIC OFF if no keys are pressed within a programmed time
period in RUN mode.
The automatic off can either be set to OFF or programmed up to 24 hours in
1-hour intervals in the SETUP MENU STANDARD screen.
Follow the steps below:
1. Press
d to move to the SETUP MENU STANDARD screen.
Press
f to select.
2. Press
d to move to the AUTOMATIC OFF screen.
The current status is displayed at the bottom
left of the screen.
3. Press or scroll
a to increase or s to decrease the duration for the
automatic off in 1-hour intervals.
4. If you would like to set the automatic off function to OFF, press or scroll
s
until 0 hours and OFF appears on the display.
5. Press
f to save and exit.
▲
!
Discuss using the automatic off function with your doctor
or healthcare team.