Specifications
This section summarizes the specifications of the R Series End Tidal Carbon Dioxide (EtCO
2
)
option.
CAPNOSTAT 5® CO
2
Sensor
Transducer Type
Mainstream
Principle of Operation
Non-Dispersive Infrared (NDIR) single beam optics,
dual wavelength, no moving parts.
Warm Up Time
Full specifications within 2 minutes at an ambient
temperature of 25º C. Capnogram in 15 seconds.
EtCO
2
Measurement Range
0 - 150 mmHg
0 - 20%
0 - 20 kPa
EtCO
2
Accuracy
(at 760 mmHg, ambient
temperature of 25°C)
0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
101 - 150 mmHg, ±10% of reading
(At respiration rates > 80 breaths per minute, all
ranges are ±12% of actual.)
EtCO
2
Resolution
0.1 mmHg 0 - 69 mmHg
0.25 mmHg 70 - 150 mmHg
EtCO
2
Stability
Short-Term Drift: Drift over four hours < 0.8 mmHg.
Long-Term Drift: Accuracy specification will be
maintained over a 120 hour period after zeroing.
EtCO
2
Noise
RMS noise of the sensor < 0.25 mmHg at
7.5% CO
2
.
EtCO
2
Rise Time (10-90%)
< 60 ms (Adult/pediatric adapters)
< 60 ms (Infant/pediatric adapters)
Respiration Rate Range
2 - 150 breaths per minute
Respiration Rate Accuracy
±1 breath per minute
Sample Flow Rate
N/A
Compensations
Barometric pressure 400 - 850 mmHg (automatic).
Operator selectable O
2
/N
2
O compensation.
EtCO
2
Alarm Limits
User selectable, High 5 - 100 mmHg, Low
0 - 95 mmHg, OFF.
Respiration Rate (RR) Alarm Limits
User selectable, High 5 - 150 respirations per
minute, Low 0 - 100 respirations per minute, OFF.
Halogenated Agents
Specification allows for halogenated anesthetic
agents that may be present at normal clinical levels.
The presence of Desflurane in the exhaled breath
beyond normal values (5-6%) may positively bias
carbon dioxide values by up to an additional 2-3
mmHg.
Airway Adapter Deadspace
Adult < 5 cc
Infant < 1.0 cc