beautypg.com

Probe adjustment – Teledyne LeCroy PP006A User Manual

Page 2

background image

Probe Adjustment

LF Adjustment

LF compensation is made by connecting the probe to the cal signal and adjusting the

compensation trimmer in the BNC-box at end of the probe. Connect the BNC end of the

probe to a channel of the oscilloscope. Connect the sprung hook end (1) and the ground lead

(2) to the CAL out and GND on the scope Calibrator area. The CAL signal is available in

many speeds. Select the Utilities in the scope to set the CAL to 1 kHz. Turn the scope

channel on. Press Auto Setup and adjust the time base and volts/div settings until the signal

on the screen contains two cycles (see figures below). Use the Trimmer tool (6) to adjust the

probe response to match the correct trim shown below.

[Note] Do not apply excess pressure to the screwdriver when you use it. The tip of

screwdriver are is easy to brake, because of made by plastics.

Undershoot

Correct

Overshoot

Typical Voltage Derating Curve

Typical Input Impedance

1.0E +02

1.0E +03

1.0E +04

1.0E +05

1.0E +06

1.0E +07

1.0E +08

1.0E +00

1.0E +02

1.0E +04

1.0E +06

1.0E +08

1.0E +10

Frequency[H z]

Zin[ohms

]

1

10

100

1000

0.1

1

10

100

1000

Frequency[M H z]

Voltage

[V

(DC+AC peak

)

]