Teledyne LeCroy PPE 2 kV User Manual

Oltage

background image

H

IGH

V

OLTAGE

P P E 2 k V

G

ENERAL

C

HARACTERISTICS

Operating temperature: 0 °C to 50 °C
Cable length: 2 m
Weight of probe + accessories: 70 gr
Coupling: 1 M

Ω DC/AC

Warranty: One year
Overvoltage category: CAT I
Pollution degree: 1

C

LEANING

The exterior of the probe only should be
cleaned with a soft cloth moistened with
either water or isopropyl alcohol.

U

SING AND

M

AINTENANCE

The probe user should note that the protection
provided by the equipment may be impaired,
if the equipment is used in a manner not
specified by the manufacturer.
To guarantee accurate performance
characteristics, machanical shocks should be
avoided, as well as damage to the cable
through excessive bending. Handle with care,
especially when the extra thin spring-tip is
fitted. All maintenance should be referred to
qualified personnel.

S

ALES

& S

ERVICE

L

E

C

ROY

Corporate Headquarters
U.S.A.: Chestnut Ridge
Phone (1) 914 578 6020
Fax (1) 914 578 5985

European Headquarters
Germany: Heidelberg
Phone (49) 6221 827 00
Fax (49) 6221 834 655

S

AFETY

I

NFORMATION

The probe is designed to be safe under the
following conditions:

Indoor use
Altitude to 2000 m
Temperature range from 0 °C to 50 °C
Maximum relative humidity 80 % for tem-
peratures up to 31 °C, decreasing linearly
to 40 % relative humidity at 50 °C

W

ARNING

Do not disassemble or disconnect probes while
they are connected to a voltage source. Use only
grounded oscilloscopes.

Attention: Refer to this manual before
using the probe.

Danger. High Voltage.

10

100

1000

10000

0,1

1

10

100

1000

Frequency [MHz]

V

o

lt

age [

V

]

E

LECTRICAL

C

HARACTERISTICS

System attenuation: 100:1
System input resistance: 50 M

Input capacitance: < 6 pF
Compensation range: 10 pF to 50 pF
System BW (-3 dB): 400 MHz typical
Max. non-destructive input voltage: 2 kV
See voltage derating curve

Impulse

DC + peak AC