Lf p, Hf p – Teledyne LeCroy PPE 20 kV User Manual

Page 2

background image

LF P

ROBE

A

DJUSTMENT

(1000:1)

Connect the probe to a 1 kHz square wave
signal. Adjust compensation trimmer in the
BNC-box for optimum square wave response.

Undershoot

Correct

Overshoot

HF P

ROBE

A

DJUSTMENT

If necessary the probe can be adjusted for high
frequency. Connect the probe to a 1 MHz square
wave signal and adjust the trimme rin order to
receive an optimum square wave response.

Undershoot

Correct

Overshoot

S

TANDARD ACCESSORIES SUPPLIED

WITH THE PROBE

Accessories

Quantities

1. Trimming tool

1

2. Ground lead with a crocodile clip

1

3. Hook

1

H

IGH

V

OLTAGE

P

ROBE

K

IT

P

K

104

Accessories

Quantities

1. Trimming tool

1

2. Ground lead with a crocodile clip

1

3. Hook

1

4. BNC Adapter

1

1

3

4

2

Derating PPE 20 kV

1000

10000

100000

0,01

0,1

1

10

100

Frequency [MHz]

Maximum Impuls Derating - PPE20kV

0

5000

10000

15000

20000

25000

30000

35000

40000

45000

10

100

1000

Duration [msec]

10%

25%

50%

100%

Duty Cycles