System performance summaries, Measurement port characteristics – Agilent Technologies 8753ET User Manual
Page 8

8
Agilent 8753ES (75-ohm systems)
Type-F test ports
Configuration
Network analyzer
8753E Option 075
Calibration kit
85039B
Test-port cables
11857B
System dynamic range
30 kHz to 50 kHz
68 dB
1
50 kHz to 300 kHz
90 dB
1
300 kHz to 16 MHz
96 dB
16 MHz to 3 GHz
106 dB
Measurement Port Characteristics
Data is shown for a Type-F female reflection port and
a Type-F male transmission port.
Frequency Range
Corrected
2
30 kHz–300 kHz
300 kHz–1.3 GHz 1.3 GHz–3 GHz
Directivity
45 dB
45 dB
40 dB
Source match
40 dB
40 dB
30 dB
Load match
45 dB
45 dB
40 dB
Reflection
±(0.060 dB
±(0.060 dB
±(0.024 dB
tracking
+0.02 dB/°C)
+0.01 dB/°C)
+0.02 dB/°C)
Transmission
±(0.033 dB
±(0.019 dB
±(0.057 dB
tracking
+0.02 dB/°C)
+0.01 dB/°C)
+0.02 dB/°C)
System Performance Summaries
1. Typical below 300 kHz.
2. Typical performance.
Transmission uncertainty
Magnitude
Phase
Magnitude
Phase
Reflection uncertainty