Measurement port characteristics – Agilent Technologies 8753ET User Manual
Page 7

Agilent 8753ES (75-Ohm Systems)
Type-N test ports
Configuration
Network analyzer
8753ES Option 075
Calibration kit
85036B
Test-port cables
11857B
System Dynamic Range
30 kHz to 50 kHz
68 dB
1
50 kHz to 300 kHz
90 dB
1
300 kHz to 16 MHz
96 dB
16 MHz to 3 GHz
106 dB
Measurement Port Characteristics
Frequency Range
Corrected
30 kHz–300 kHz
2
300 kHz–1.3 GHz 1.3 GHz–3 GHz
Directivity
48 dB
48 dB
43 dB
Source match
47 dB
41 dB
35 dB
Load match
48 dB
48 dB
43 dB
Reflection
±(0.004 dB
±(0.010 dB
±(0.019 dB
tracking
+0.02 dB/°C)
+0.01 dB/°C)
+0.02 dB/°C)
Transmission
±(0.018 dB
±(0.016 dB
±(0.033 dB
tracking
+0.02 dB/°C)
+0.01 dB/°C)
+0.02 dB/°C)
Uncorrected
2
Directivity
20 dB*
35 dB
30 dB
Source match
16 dB**
16 dB
16 dB
Load match
15 dB**
18 dB
16 dB
Reflection
±2.5 dB
±1 dB
±1 dB
tracking
Transmission
±2.5 dB
±1 dB
±1 dB
tracking
Crosstalk
90 dB***
100 dB
100 dB
*15 dB, 30 to 50 kHz
**10 dB, 30 to 50 kHz
***60 dB, 30 to 50 kHz
7
1. Typical below 300 kHz.
2. Typical performance.
Transmission uncertainty
Magnitude
Phase
Magnitude
Phase
Reflection uncertainty