Agilent 8753ES (50-ohm systems)
7-mm test ports
Configuration
Network analyzer
8753ES
Standard and Option 006
Calibration kit
85031B
Test-port cables
11857D
System dynamic range
30 kHz to 50 kHz
70 dB
1
50 kHz to 300 kHz
90 dB
1
300 kHz to 16 MHz
100 dB
16 MHz to 3 GHz
110 dB
Option 014
108 dB
3 GHz to 6 GHz
105 dB
Option 014
103 dB
Measurement Port Characteristics
Frequency Range
Corrected
30 kHz–
300 kHz–
1.3 GHz–
3 GHz–
300 kHz
2
1.3 GHz
3 GHz
6 GHz
Directivity
55 dB
55 dB
51 dB
46 dB
Source match 55 dB
51 dB
49 dB
43 dB
Load match
55 dB
55 dB
51 dB
46 dB
Reflection
±(0.001 dB
±(0.001 dB
±(0.005 dB
±(0.020 dB
tracking
+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)
Transmission ±(0.008 dB
±(0.006 dB
±(0.009 dB
±(0.021 dB
tracking
+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)
Uncorrected
Directivity
20 dB*
35 dB
30 dB
25 dB
Source match 18 dB**
16 dB
16 dB
14 dB
Load match
18 dB**
18 dB
16 dB
14 dB
Reflection
±2.5 dB
±1 dB
±1 dB
±1.5 dB
tracking
Transmission ±2.5 dB
±1 dB
±1 dB
±1.5 dB
tracking
Crosstalk
90 dB***
100 dB
100 dB
90 dB
*15 dB, 30 kHz to 50 kHz
**10 dB, 30 kHz to 50 kHz
***60 dB, 30 kHz to 50 kHz
System Performance Summaries
1. Typical below 300 kHz.
2. Typical performance.
Reflection uncertainty
Magnitude
Phase
Magnitude
Phase
Transmission uncertainty