Ransburg, Installation, Setup procedure – Ransburg MMA-303 Direct_Ind Test Stand User Manual

Page 12: Initial test settings

background image

Ransburg

SETUP PROCEDURE

Note: Refer to MMA-303 Service Manual for all

parts and part number identifications.

1. Ensure that ground connections, cart ground,

and applicator ground are made to a true earth

ground.

INSTALLATION

2. Provide air filtration to the manifold that has the

capability of passing 136 SCFM with particulate

removal of .3 to .6 micron particle size (bearing

air filtration is provided with the stand).

3. Provide a minimum 3/4-inch ID air line to the

filter at the manifold inlet. A minimum of 100 psig

pressure is required at the manifold inlet.

4. Install water supply line and air pressure line

to the pressure pot.

6. Plug the unit into a suitable 110/120V outlet.

INITIAL TEST SETTINGS

1. High voltage is off unless instructed to turn on.

Main power off.

2. Fluid enable off (toggle switch lever in the down

position).

3. Attach the applicator to the Test Stand bracket

with three (3) M10 screws (78268-00).

4. Connect tubing from the Test Stand to the ap-

plicator as labeled.

5. Supply 100 psig air pressure to the stand, refer-

ence supply air gage on panel. If required, adjust

air input regulator attached to stand.

6. Set speed select to 00.

7. Adjust bearing air regulator to 90 psig minimum.

5. Fill the pressure pot liner with approximately

one (1) gallon of either de-ionized or distilled

water.

N O T E

†

When opening dump valve for any

test, hold tubing securely by hand or

clamp to prevent whipping.

MMA-303 Test Station - Installation

9

LN-9255-06.2

†

If true earth ground is not made, injury to

personnel or serious damage may occur to

the equipment.

†

Ensure ALL items in the test area are

grounded. Insure there are NO open contain-

ers of flammable materials in the test area.

W A R N I N G

!

†

NEVER use solvent as a test fluid for this

device!

†

Using any material OTHER than distilled

water or de-ionized water may cause fire dur-

ing the test. Use only distilled water or de-

ionized to test the applicator.

W A R N I N G

!