4.2. IR ADAPTOR


The IR Adaptor takes advantage of a multi-photon absorption process to extend the sensitivity of the Beamage-
3.0
beam profiling camera to a portion of the near-IR spectrum.

The module converts wavelengths ranging between 1495 nm and 1595 nm (telecom wavelengths band) to
shorter wavelengths ranging between 950 nm and 1075 nm.

After going through an anti-reflection coated input window, the laser beam gets instantaneously converted with
high resolution, low distortion and good uniformity.

The IR Adaptor can be C-mounted onto the entrance port of the Beamage-3.0 camera.

Figure 10: IR Adaptor

Table 9: IR Adaptor Specifications

Active Area

27.5 mm Ø

IR Spectral Range

1495 nm

– 1595 nm

Peak IR Sensitivity

1510 nm and 1540 nm

Converted Wavelengths

950 nm

– 1075 nm

Pixel Multiplication Factor

3.29

Minimum Beam Size

230 µm

Maximum Beam Size

19 mm

Maximum Resolution

12 lp/mm over active area
40 lp/mm at sensor focal plane

Distortion

-1.0% barrel distortion (inverted image)

Linearity

Non-Linear, IR converted output

∝ IR input

intensity ^1.41

Spectral Transmission

360 nm

– 2000 nm @ F30.8

Damage Threshold

1 W/cm

2

Dimensions

46 mm Ø x 97 mm L

Operating Temperature

-10°C to +40°C

Weight

210 g

Part Number

201061