Section 2: Calibration

2-1 NOTES

Before beginning the calibration procedure, be sure to power off the scale and
empty the weighing platform.

Use calibration standard of more than one-sixth (

1

/

6

) of the scale’s

maximum capacity.

During calibration, use standard weights that are not over the scale’s
maximum capacity. For example, if calibrating a 25-pound capacity scale,
use a standard 5-pound weight to calibrate it.

2-2 KEYPAD FUNCTION CHART

The SET key is used to confirm a setting.


The CHANGE key is used to move through the settings.


The ZERO key is used to increase the number in the display.

1. Turn the power OFF.


2. Press

the

SET key and hold while turning

the scale ON by pressing the OFF-ON key.

3. The screen will display the number 1.


4. Press

the

ZERO key to display the number 2.

5. Press

the

SET key to display the number 3.

SET

CHANGE

ZERO

+

SET

1

OFF ON

OFF ON

2

ZERO

( + )

3

SET