SECTION 3: USING THE SCALE
3.1. POWER- ON:
Because the scale uses special low-power circuitry, no warm-up time is required. Weight
readings will be accurate as soon as the unit is powered on and set to zero.
When the unit is powered, the liquid crystal display (LCD) will show the software part
number and the revision followed by either "0.00", or "------". Dashes are displayed to
indicate the scale is registering a weight upon startup. Press the [
ZERO
] key to set the
display to “0.00” and start weighing.
3.2. KEYS:
The keys for operating the scale are located on the main display and on the remote
display.
1. Pressing the [
ZERO
] key resets the display to indicate zero (0).
a. The zero range is set at 2% or 3.00 lbs when set for Canadian use
b. The zero range is set at 100 % or 150 lbs when set for USA use
2. The [
ZERO
] key function will be inhibited if the instrument detects any of the
following conditions:
a. Motion on the platform
b. An underload condition
c. An overload condition
d. Outside of programmed zero range
3. Pressing the [
UNITS
] key toggles the weighing units and the display 'indicators'
between "lb" to "kg". Verify the units you want to use by noting the arrow 'indicator'
on the display.