SECTION 4: CUSTOMER CARE

4.1. CLEANING:

Clean by wiping the scale assembly off with a damp cloth only. Do not use running water,

harsh chemicals, or allow liquids to drip onto the scale/display.

4.2. OPERATOR INSTRUMENT PROMPTS:

Prompt

Description

HiCAP

Load 150 lbs/ 68 kg or greater - over capacity.

LoCAP

Scale is below normal Zero range - under capacity.

“ - - - - - - ”

Shown at startup. Indicates the scale is not within the center-of-zero

range. Press the [

ZERO

] key to go to weigh mode.

“ - - - - ”

Shown at startup. Motion is preventing the scale from entering the
weigh mode



4.3. TROUBLESHOOTING

- In the event the scale does not function properly, check the

following:

Problem

Possible Source / Remedy

No Display

Power OFF, plug disconnected, cord damage,
faulty USB port or AC adapter. IF you are using a remote display, check
the main display first, if it's OK then check the cable plug connection on
the remote.
Unplug then plug in the power cord to reset the program.

Incorrect Weight

Check platform for binding or rubbing, reposition scale so all sides are
clear. Ensure correct UNITS are displayed (lb or kg). For scales in counter
inserts, make sure there is nothing jammed around edges Remove load,
press the [

ZERO

] key to set the scale to “0.00”, then reweigh.

[

ZERO

] key will NOT reset zero

Motion on the platform, ensure that the platform is empty. Check platform
for binding or rubbing, reposition scale so all sides are clear. For scales in
counter inserts, make sure there is nothing jammed around edges.
Unplug then plug in the power cord to reset the program