beautypg.com

Teledyne TED-200 - Medical application oxygen monitor User Manual

Page 22

background image

3 Maintenance and Troubleshooting

Teledyne Analytical Instruments

3-2

If the unit is left on after this, the LCD display will eventually shut down, the red LED

will stay lit, and the beeper will emit a continuous tone. If the batteries are still not replaced,
the LED and beeper will drain the remaining power and go completely dead.

If the TED 200-T will not be used for a period of 30 days or more, the batteries should

be removed prior to storage.

To replace batteries, see chapter 2, section 2.1.1. Remember, the unit must be calibrated

after the batteries are replaced.

3.3

Sensor Maintenance

DO: check the sensor for damage or leaks before use.

DO: recalibrate after replacing the sensor.

DON’T: immerse the T-7 sensor in liquid.

DON’T: autoclave the T-7 sensor.

DON’T: open or try to repair the sensor.

Before every use, the sensor, cable and connections should be checked. Check the

sensor for leaks and condensation. Check the cable for splitting or cracked insulation. Make
sure the connections are tight and dry.

In the event that the sensor has been damaged, consult the Material Safety Data Sheet in

the Appendix for handling guidelines.

To replace the sensor, see chapter 2, section 2.1.2.

3.4

Troubleshooting

The TED 200-T provides a variety of built-in safety features which prevent instrument

use when a fault or incorrect calibration procedure is detected. When a unit repeatedly
displays the messages REPLACE SENSOR or SEE TROUBLESHOOT, sounds/flashes
the alarms continuously, or refuses to turn on, it may be an indication of an erroneous cali-
bration, an intermittent or faulty connection, an expired sensor, or a low battery. To deter-
mine where the difficulty lies, refer to the following guidelines.

NOTE: The TED 200T Service Manual (P/N SMC41901.02) is available from the

factory and contains circuit diagrams, test procedures and other informa-
tion which will assist qualified technical personnel in repairing the Monitor.