Teledyne TED-200 - Medical application oxygen monitor User Manual
Page 19

Operations 2
2-9
Teledyne Analytical Instruments
2.3
Do’s and Don’ts
– DO –
•
Read all of the directions before using for the first time.
•
Calibrate every 8 hours or before every use.
•
Visually inspect the sensor for leakage before each use.
•
Use 2-point calibration every time: 21%, then 100%.
•
Check the HI and LO alarm settings prior to each use.
•
Test batteries regularly (replace immediately when indicated).
•
Keep the unit, sensor and connections dry, or on the dry side of the breathing
circuit.
•
Recalibrate after replacing the batteries.
•
Recalibrate after replacing the sensor.
•
Use properly installed alkaline batteries only.
•
Make sure the T-7 sensor is properly attached.
•
Use the plastic flow diverter only when using the tee adapter.
•
Remove and save the plastic flow diverter when using the unit in chamber
applications (incubators, tents, etc.)
•
Perform an alarm test before each use (press the ALARM TEST key twice).
•
Clean the case with isopropyl alcohol or mild detergent only.
– DON’T –
•
Use the TED 200-T if you suspect any malfunction.
•
Use the TED 200-T in the presence of flammable gases.
•
Use anything but alkaline batteries.
•
Autoclave or freeze the T-7 sensor or TED 200-T unit.
•
Open or try to repair a leaking or broken sensor.
•
Immerse the unit or sensor in any liquid.
•
Pass hot or cold gas mixtures over the sensor.
•
Expose the unit to devices that produce high levels of radio, short wave,
microwave, x-ray, or high frequency interference.
•
Use cleaning agents or liquids in the cable receptacles or around the battery
compartment.
•
Place the TED 200-T unit itself in a water vapor-saturated environment.
•
Expose the LCD to excessive sunlight.
•
Expose the TED 200-T to a condensing water environment such as a mist
tent.