Teledyne TED-60 - Medical application oxygen monitor User Manual
Page 16
2 Operations
2 Operations
2 Operations
2 Operations
2 Operations
Teledyne Analytical Instruments
2-8
DO:
DO:
DO:
DO:
DO:
•
Read all of the directions before using for the first time.
•
Calibrate every 8 hours or before every use.
•
Keep the unit, sensor and connections dry, or on the dry side of the
breathing circuit.
•
Recalibrate after replacing the batteries.
•
Recalibrate after replacing the sensor.
•
Use properly installed alkaline
alkaline
alkaline
alkaline
alkaline batteries only.
•
Make sure the T-7 sensor is properly attached.
•
Visually inspect the sensor for leakage or water condensation on the
sensing surface before each use.
•
Use the plastic flow diverter only when using the tee adapter.
•
Remove and save the plastic flow diverter when using the unit in cham-
ber applications (incubators, tents, etc.)
•
Clean the case with isopropyl alcohol only, or in extreme cases, a mild
detergent.
•
Remove the batteries prior to storage.
DON’T:
DON’T:
DON’T:
DON’T:
DON’T:
•
Use the TED 60T if you suspect any malfunction.
•
Use the TED 60T in the presence of flammable gases.
•
Use anything but alkaline batteries.
•
Autoclave or freeze the T-7 sensor or TED 60T unit.
•
Open or try to repair a leaking or broken sensor.
•
Immerse the unit or sensor in any liquid.
•
Pass hot or cold gas mixtures over the sensor.
•
Expose the unit to devices that produce high levels of radio, short wave,
microwave, x-ray, or high frequency interference.
•
Use cleaning agents or liquids in the cable receptacles or around the
battery compartment.
•
Place the TED 60T unit itself in a water vapor-saturated environment.
•
Expose the LCD to excessive sunlight.
•
Expose the TED 60T to a condensing water environment such as a mist
tent.
Do’s and Don’ts
Do’s and Don’ts
Do’s and Don’ts
Do’s and Don’ts
Do’s and Don’ts