Discrepancy in Readings

Discrepancy in Readings

Discrepancy in Readings

Discrepancy in Readings

Discrepancy in Readings

The TED 60T is intended to be used as a secondary

secondary

secondary

secondary

secondary means to verify the accuracy

of and check the oxygen concentration leaving another oxygen mixing device or pri-
mary life support system (i.e., a blender or respirator). Whenever there is a significant
difference in the oxygen readings between the primary support system and the analyzer,
the discrepancy must be resolved immediately. The information obtained from the TED
60T should never be used to make adjustments to the primary life-support system, but
should only be used as an indication that the primary device may

may

may

may

may require service and/or

calibration.