Installation Insta-Trans

Teledyne Analytical Instruments

14

1. Insert the tube into the tube fitting, and finger-tighten the nut

until the tubing in the fitting cannot be rotated freely by hand.
(This may require an additional 1/8 turn beyond finger-tight.)

2. Hold the fitting body steady with a backup wrench, and with

another wrench rotate the nut another 1-1/4 turns.

Figure 3-1: Insta-Trans Connections and Mounting Dimensions

SAMPLE IN: Gas connections are made at the

SAMPLE IN

and

EXHAUST OUT

connections.

The gas pressure in should be reasonably regulated. A flow control

device must be installed before the sample in port if the sample pressure
is above 0.5 psig.