beautypg.com

Operational theory – Teledyne Insta Trans - trace and percent oxygen transmitter User Manual

Page 13

background image

Trace Oxygen Transmitter

Operational Theory

Teledyne

Analytical

Instruments

5

Operational Theory

2.1 Introduction

The transmitter is composed of three subsystems:

1. Micro-fuel Cell Sensor

2. Sample System

3. Electronic Signal Processing, Display and Control

The sample system is designed to accept the sample gas and

transport it through the transmitter without contaminating or altering the
sample prior to analysis. The Micro-fuel Cell is an electrochemical
galvanic device that translates the amount of oxygen present in the
sample into an electrical current. The electronic signal processing,
display and control circuits simplify operation of the transmitter and
accurately displays the sampled data.

2.2 Micro-fuel Cell Sensor

2.2.1 Principles of Operation

The oxygen sensor used in the Insta-Trans series is a Micro-fuel

Cell designed and manufactured by Analytical Instruments. It is a sealed
plastic disposable electrochemical transducer.

The active components of the Micro-fuel Cell are a cathode, an

anode, and the 15% aqueous KOH electrolyte in which they are
immersed. The cell converts the energy from a chemical reaction into an
electrical current in an external electrical circuit. Its action is similar to
that of a battery.

There is, however, an important difference in the operation of a

battery as compared to the Micro-fuel Cell: In the battery, all reactants
are stored within the cell, whereas in the Micro-fuel Cell, one of the
reactants (oxygen) comes from outside the device as a constituent of the
sample gas being analyzed. The Micro-fuel Cell is therefore a hybrid

This manual is related to the following products: