Appendix model 9700 – Teledyne 9700 - Flue gas analysis system User Manual

Page 46

background image

A-2

Appendix

Model 9700

TELEDYNE BROWN ENGINEERING

Analytical Instruments

OXYGEN ANALYSIS SECTION

Ranges: 0–5%, 0–10%, 0–25% O

2

Sensitivity: 0.5% of full scale

Accuracy: +/- 2% of full scale at const. temp.

+/- 5% of full scale across temp. range

Response Time: 90% of O

2

in 45–60 seconds

Calibration: Air

Cooling Water: 10–100 psi,

1

/

4

–1

1

/

2

gpm

Sensor: Class A-3 Micro-Fuel Cell

(warranted for 6 months in normal flue gas
applications)

Alarms (optional): 1 or 2 (3 Amp resistive)

COMBUSTIBLE GAS SECTION

Ranges: 0–5%, combustibles (CH

4

equiv. standard)

Sensitivity: 0.5% of full scale

Accuracy: +/- 2% of full scale

Response Time: 90% of combustibles in 15 seconds

Calibration: With span gas composed of 4.0–4.8%

methane content (balance nitrogen)

Air Supply: Compressed air regulated to 5–25 psig,

2 SCFH typical required consumption

Sensor: Low temperature catalytic bead

(warranted for 6 months)

Alarms Contacts: 1 SPDT (3 Amp resistive) non-latching

OPTIONAL CALIBRATION KIT

# 6 size (G-4l) cylinder of span gas
(4–4.8% methane, bal. nitrogen)

Pressure and flow regulators

Adapter: 5 ft. (152.4 cm) of 3/16 "
(4.74 mm) ID silicone tubing