Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 91

UV Fluorescence SO2 Analyzer
Appendix A
Teledyne
Analytical
Instruments
79
Appendix A-6: Terminal Command Designators,
Revision C.0
Table A-7: Terminal Command Designators, Revision C.0
COMMAND
ADDITIONAL COMMAND SYNTAX
DESCRIPTION
? [ID]
Display help screen and this list of
commands
LOGON [ID]
password
Establish connection to instrument
LOGOFF [ID]
Terminate connection to instrument
T [ID]
SET ALL| name| hexmask
Display test(s)
LIST [ALL| name | hexmask] [NAMES|HEX]
Print test(s) to screen
name
Print single test
CLEAR ALL| name | hexmask
Disable test(s)
W [ID]
SET ALL| name| hexmask
Display warning(s)
LIST [ALL| name | hexmask] [NAMES|HEX]
Print warning(s)
name Clear
single
warning
CLEAR ALL| name | hexmask
Clear warning(s)
C [ID]
ZERO|LOWSPAN|SPAN [1|2]
Enter calibration mode
ASEQ number
Execute automatic sequence
COMPUTE ZERO|SPAN
Compute new slope/offset
EXIT Exit
calibration
mode
ABORT
Abort calibration sequence
D [ID]
LIST
Print all I/O signals
name[=value]
Examine or set I/O signal
LIST NAMES
Print names of all diagnostic tests
ENTER name
Execute diagnostic test
EXIT Exit
diagnostic
test
RESET [DATA] [CONFIG] [exitcode] Reset
instrument
PRINT ["name"] [SCRIPT]
Print iDAS configuration
RECORDS ["name"]
Print number of iDAS records