background image

UV Fluorescence SO2 Analyzer

Appendix A

Teledyne

Analytical

Instruments

79

Appendix A-6: Terminal Command Designators,
Revision C.0

Table A-7: Terminal Command Designators, Revision C.0

COMMAND

ADDITIONAL COMMAND SYNTAX

DESCRIPTION

? [ID]

Display help screen and this list of
commands

LOGON [ID]

password

Establish connection to instrument

LOGOFF [ID]

Terminate connection to instrument

T [ID]

SET ALL| name| hexmask

Display test(s)

LIST [ALL| name | hexmask] [NAMES|HEX]

Print test(s) to screen

name

Print single test

CLEAR ALL| name | hexmask

Disable test(s)

W [ID]

SET ALL| name| hexmask

Display warning(s)

LIST [ALL| name | hexmask] [NAMES|HEX]

Print warning(s)

name Clear

single

warning

CLEAR ALL| name | hexmask

Clear warning(s)

C [ID]

ZERO|LOWSPAN|SPAN [1|2]

Enter calibration mode

ASEQ number

Execute automatic sequence

COMPUTE ZERO|SPAN

Compute new slope/offset

EXIT Exit

calibration

mode

ABORT

Abort calibration sequence

D [ID]

LIST

Print all I/O signals

name[=value]

Examine or set I/O signal

LIST NAMES

Print names of all diagnostic tests

ENTER name

Execute diagnostic test

EXIT Exit

diagnostic

test

RESET [DATA] [CONFIG] [exitcode] Reset

instrument

PRINT ["name"] [SCRIPT]

Print iDAS configuration

RECORDS ["name"]

Print number of iDAS records