UV Fluorescence SO2 Analyzer

Appendix A

Teledyne

Analytical

Instruments

73

SIGNAL NAME

BIT OR CHANNEL

NUMBER

DESCRIPTION

0–7

Spare

Control outputs, U21, J1008, pins 9–12 = bits 0–3, default I/O address 325 hex

0–3

Spare

Alarm outputs, U21, J1009, pins 1–12 = bits 4–7, default I/O address 325 hex

ST SYSTEM OK2

4

1 = system OK

0 = any alarm condition or in diagnostics
mode

5–7

Spare

A status outputs, U24, J1017, pins 1–8 = bits 0–7, default I/O address 323 hex

ST_SYSTEM_OK 0

0 = system OK

1 = any alarm condition

ST_CONC_VALID 1

0 = conc. valid

1 = warnings or other conditions that
affect validity of concentration

ST_HIGH_RANGE 2

0 = high auto-range in use

1 = low auto-range

ST_ZERO_CAL 3

0 = in zero calibration

1 = not in zero

ST_SPAN_CAL 4

0 = in span calibration

1 = not in span

ST_DIAG_MODE 5

0 = in diagnostic mode

1 = not in diagnostic mode

ST_LOW_SPAN_CAL 6

0 = in low span calibration

1 = not in low span

7 Spare

B status outputs, U27, J1018, pins 1–8 = bits 0–7, default I/O address 324 hex

ST_LAMP_ALARM 0

0 = lamp intensity low

1 = lamp intensity OK

ST_DARK_CAL_ALARM 1

0 = dark cal. warning

1 = dark cal. OK

ST_FLOW_ALARM 2

0 = any flow alarm

1 = all flows OK

ST_PRESS_ALARM

3

0 = any pressure alarm