Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

13

The RS-232 protocol allows some flexibility in implementation in

the choice of values for certain parameters. Table 3-2 lists the RS-232
values required by the Model 3300ZB implementation.

Table 3-2: Required RS-232 Options

Parameter Setting

Baud

2400

Byte

8

bits

Parity

none

Stop

Bits

1

Message Rate

2 per second

3.4 Gas Connections

The standard Model 3300ZB has two inlet (sample and span) and

one outlet port. All connections have 1/8” tube fittings installed. A
switching valve is provided to feed the unit with either sample or
calibration gas as shown in Figure 3-3.

A flowmeter and flow controller valve are part of the system and

will assist in setting the flow of the gas.

Sample flow should be adjusted to 2 SCFH. The sample vent

connection should not restrict the sample flow. The sensor is designed to
operate at atmospheric pressure. Restricting the sample vent line will
result in pressurizing the sensor and altering the O

2

reading.

The pressure required will depend on the sampling system. When

venting into a constant pressure, such as the atmosphere, controlling
input pressure is simple. If you are venting into a system of varying
pressure, then some form of pressure regulation is required to maintain a
constant pressure.

Sample and span pressure must NOT exceed 50 psig (345 Kpa).

Figure 3-3 shows a standard piping diagram for positive pressure

service.