beautypg.com

Teledyne 326, 327 and 328 - Oxygen analyzers User Manual

Page 32

background image

Teledyne Analytical Instruments

4-4

P

P

P

P

Per

er

er

er

ercent Oxyg

cent Oxyg

cent Oxyg

cent Oxyg

cent Oxygen Anal

en Anal

en Anal

en Anal

en Analyz

yz

yz

yz

yzer

er

er

er

er

Section 4

Section 4

Section 4

Section 4

Section 4

5) Connect analyzer input to the sample gas and retrim the throttle

valve for an indicated flowrate of 2.0 SCFH, or insert probe
holder back into its holder.

Note:

Note:

Note:

Note:

Note:

Analyzer accuracy is not affected by variations in sample flowrate. The

Analyzer accuracy is not affected by variations in sample flowrate. The

Analyzer accuracy is not affected by variations in sample flowrate. The

Analyzer accuracy is not affected by variations in sample flowrate. The

Analyzer accuracy is not affected by variations in sample flowrate. The
recommended flowrate of 2.0 SCFH insures maximum speed or

recommended flowrate of 2.0 SCFH insures maximum speed or

recommended flowrate of 2.0 SCFH insures maximum speed or

recommended flowrate of 2.0 SCFH insures maximum speed or

recommended flowrate of 2.0 SCFH insures maximum speed or
response.

response.

response.

response.

response.

6) Select the range of analysis that will provide the best possible

resolution of the oxygen content of the sample gas.

7) Allow sample gas, not calibration air, to flow through the

analyzer. For best results, let sample gas flow through the
analyzer for 24 hours. This allows the Micro-Fuel Cell to
equilibrate.

Important:

Important:

Important:

Important:

Important:

It is particularly vital to the condition of the A-3 cell employed in

It is particularly vital to the condition of the A-3 cell employed in

It is particularly vital to the condition of the A-3 cell employed in

It is particularly vital to the condition of the A-3 cell employed in

It is particularly vital to the condition of the A-3 cell employed in
flue gas (or high CO

flue gas (or high CO

flue gas (or high CO

flue gas (or high CO

flue gas (or high CO

2

2

2

2

2

-content applications) that the cell not be

-content applications) that the cell not be

-content applications) that the cell not be

-content applications) that the cell not be

-content applications) that the cell not be

exposed to a CO

exposed to a CO

exposed to a CO

exposed to a CO

exposed to a CO

2

2

2

2

2

-free environment for a prolonged period of

-free environment for a prolonged period of

-free environment for a prolonged period of

-free environment for a prolonged period of

-free environment for a prolonged period of

time.

time.

time.

time.

time.

4.6 Operational Calibration

4.6 Operational Calibration

4.6 Operational Calibration

4.6 Operational Calibration

4.6 Operational Calibration

After the 24-hour equilibration period following the installation of any

new cell, or whenever it is desirable to recheck the calibration of the instru-
ment, use the following calibration procedure:

1) Place the range selector switch on the 0-25% position.

2) Connect the analyzer input to a supply of calibration (span) air

and adjust the flowrate to 2.0 SCFH.

3) Unlock and adjust the span control until the meter pointer is in

precise coincidence with the scale calibration mark and relock the
span control.

4) Return the instrument to service by restoring sample flow at the

proper rate (2.0 SCFH), and selecting the range that provides the
best possible resolution of the sample oxygen.

4.7 Calibration of Alarm Equipped Instruments

4.7 Calibration of Alarm Equipped Instruments

4.7 Calibration of Alarm Equipped Instruments

4.7 Calibration of Alarm Equipped Instruments

4.7 Calibration of Alarm Equipped Instruments