Precautions

1. Read this chapter in its entirety before installing the unit.

2. Upon receipt of the instrument, inspect the entire unit for

damage. Notify the shipper if any damage is found. Check the
unit, probe, and included accessories for broken or loose parts.

3. The sample must be free of entrained solids and water. A high

humidity sample is ideal since it will prevent water loss from
the cell electrolyte.

4. The Micro-Fuel Cell sensor electrolyte is caustic. Do not

attempt to open or repair the sensor. Leaking or exhausted
sensors should be disposed of in accordance with local
regulations. Please refer to the Material Safety Data Sheet in the
Appendix to learn about potential hazards and corrective action
in case of accident.

5. Do not scratch, puncture, or damage the sensing membrane of

the Micro-Fuel Cell sensor. Damage to the membrane will
require replacement of the sensor.

Location & Mounting

The unit is designed to be panel-mounted in a general purpose area,

away from moisture and the elements. The unit should be installed at
viewing level in a sheltered area.

Refer to the Outline diagram (dwg. C-57490) for the physical dimen-

sions of the analyzer.