13

TELEDYNE ANALYTICAL INSTRUMENTS

329R

Appendix

Appendix

Specifications

Ranges:

Standard: 0-3% and 0-10% O

2

with

0.01% O

2

resolution over full scale.

Optional: 0-1% O

2

0-25% O

2

0-100% O

2

Air Calibration Range:

0-25% O

2

with 0.1% O

2

resolution over

full scale.

Analog Output:

Standard: 0–10 VDC, negative ground

4-20 mA, negative ground
plus contact closures for range
identification.

Optional: 4-20 mA, isolated

0-20 mA, isolated
0-20 mA, negative ground
User specified voltage output,
isolated or neg. gr.

Display:

Digital 3½-digit LCD.

Alarms:

One high alarm, adjustable & fail-safe.
One low alarm, adjustable & fail-safe.
One low alarm, fixed at 2% full scale and
fail-safe.
Form “C” SPDT contacts rated at 5A,
230 VAC.

Accuracy:

±2% of full scale at constant temperature
±5% of full scale throughout operating
temperature range.

Response:

90% of full scale in 20 seconds.

Operating Temp. Range:

32 °F to 122 °F (0 °C to 50 °C)

Power:

85-264 VAC, 47-440 Hz or 10-36 VDC.

Enclosure:

General purpose, panel mount