TELEDYNE ANALYTICAL INSTRUMENTS

1

Trace Oxygen Analyzer Model 316RA/RB/RAD/RBD

Introduction

The Teledyne Analytical Instruments (TAI) Series 316 Trace Oxygen

Analyzer uses a unique Micro-Fuel Cell to measure the concentration of
oxygen in a gas stream. The cell has an absolute zero and produces a linear
output from the low parts per million (ppm) level through 21% (210,000
ppm) oxygen. When coupled with a “state of the art” two stage amplifier
system that incorporates operational amplifiers, the cell provides linear
trace oxygen analysis in the standard ranges of 0-10, 0-100, 0-1000, and 0-
10,000 ppm. The cell is specific for oxygen and is not influenced by
sample flow rate or the presence of hydrocarbons or halogenated hydrocar-
bons. The instrument may be calibrated with air, eliminating the need for
certified trace oxygen calibration gases.