beautypg.com

Teledyne 316RA / RB / RAD / RBD - Oxygen analyzers User Manual

Page 19

background image

TELEDYNE ANALYTICAL INSTRUMENTS

15

Trace Oxygen Analyzer Model 316RA/RB/RAD/RBD

Appendix

Specifications

Ranges:

Standard: 0-10 ppm, 0-100 ppm, 0-1,000
ppm, 0-10,000 ppm, CAL oxygen
Digital: 0-10 ppm, 0-100 ppm, 0-1,000
ppm, 1%, 25%

Sensitivity:

0.5% of full scale

Accuracy:

±2% of full scale at constant temperature
(except ±1 ppm for 0-10 ppm range)
±5% of full scale over the operating
temperature range
(except ±1 ppm for 0-10 ppm range)

Response Time:

At 77°F (25°C):
0-10 ppm, 90% in less than 45 seconds
0-100 ppm, 90% in less than 30 seconds
0-1000 ppm, 90% in less than 10 seconds
0-10,000 ppm, 90 in less than 10 seconds

Operating Temperature:

+32°F to +125°F (0°C to +52°C)

Micro-Fuel Cell Class:

B-2

Micro-Fuel Cell Warranty:

6 months

Voltage Signal Output:

0-1VDC or less (standard)

Current Signal Output:

1-5, 4-20 or 10-50 mADC (optional)

Alarm Output:

(Optional) 1 or 2 DPDT Form C relays
(3A resistive)

Alarm Deadband (Hysteresis): 0.1% of full scale

Power Requirement:

115 VAC, 50/60 Hz, 30 W (100/220
VAC optional)