APPENDIX

Specifications:

Ranges: Three decade ranges: Minimum Full scale(ppm)

20 ppm Hydrogen balance Argon

25 ppm Hydrogen balance Nitrogen, Air

100 ppm Nitrogen balance Argon, Helium
100 ppm Helium balance Air
150 ppm Nitrogen, Carbon Monoxide balance Hydrogen
200 ppm Oxygen balance Argon

Range ID: 3 range ID contacts

Local Range ID LED’S

Detector: Thermal conductivity sensor

Signal Output: 0-1VDC negative ground

4-20maDC Isolated ground

Accuracy:

±2% full scale for most binary mixture
at constant temperature. ± 5% of Full
Scale over operating temperature (once
temperature equilibrium has been achieved)

Operating

Temperature:

20 – 30°C (68° F to 85° F)

Sample
Requirement:

Sample: 0.3SCFH
Reference: 0.1SCFH

Display:

Digital LED read out (3½ Digit)

Mounting:

Flush panel mount.

Response Time:

90% of full scale less then 60 seconds.
For most application.

Stability:

Less then 2% full scale drift over 24
hours

Power

requirement:

110VAC, 50/60Hz (220VAC optional)