Teledyne Analytical Instruments

26

Thermal Conductivity Analyzer Model 212R

7.4.3 Span Gas.

Containing an equivalent impurity. Again, the span gas must be procured

from a supplier who will certify its composition. The range switch must be in
range 3 when the span gas is introduced.

IMPORTANT: The accuracy of the analysis is directly related to the

users knowledge of the span gas composition.

Reference and Sample Flowrates 0.3 SCFH

Bypass Flowrate 3.0 SCFH

NOTE: Cylinder and sample stream pressure settings must

be adjusted in order to maintain the above flowrates
when switching from one sample to another.

RECOMMENDED SPARE PARTS LIST

QUANTITY

PART NO.

DESCRIPTION

1

A-6251

Measuring Cell Block Assembly

1

A-6981

Proportional Temperature Control

1

F-45

Sample Flowmeter Tube

1

F-22

Sample Flowmeter Bypass Tube

A minimum charge applies to all spare part orders.

IMPORTANT: Orders for replacement parts should include the part

number, the model and serial number of the analyzer in which they are to be
used.