4.5 HOLD

Function

For prolonged observation of a reading, press HOLD
while in measurement mode to freeze the display.

4.

To hold a measurement, press HOLD while in
measurement mode. [HO] will appear on the
display.

5.

To release the held value, press the HOLD again.
[HO] will disappear and measure is resumed.

NOTE:

This meter shuts off automatically after 20 minutes of the last key press.

If the meter is shut off either automatically or manually, the HOLD value
will be lost.

5.

ADVANCED SETUP FUNCTIONS

5.1

Advanced Setup Overview

Advanced setup allows customization settings such as; selecting electrode’s cell
constant, normalization temperature, temperature coefficient, TDS factor (TDS 6+),
automatic or manual calibration (COND 6+), single-point or multi-point calibrations
(COND 6+ and TDS 6+), and to reset meter to factory default.

To enter advanced setup mode:

6.

Switch off the meter.

7. Press

MODE

and then ON, holding both keys for 2 seconds. Release the ON key

before releasing the MODE key.

8.

If successful, [StUP] will appear briefly followed by [CELC] (for COND 6+ and
TDS 6+) or [tCo%] (for SALT 6+).

9. Press

or to select the desired advanced setup function.

S

µ

S

µ

HOLD

ENTER

HO