beautypg.com

8 switching the meter on, Cond 6 – Nova-Tech SALT 6+ User Manual

Page 12

background image

Instruction Manual

COND 6+, TDS 6+, SALT 6+

-7-

2.8

Switching the Meter On

Press ON/OFF to power up your meter. Your meter will cycle through various setup
parameters when switched on.

1.

The first screen shows the model [Con6] [tdS6] [SAL6].

2.

The second screen shows the nominal cell constant value. The Cond 6+ and TDS
6+ meters can accept electrodes with k = 0.1, 1.0 or 10.0 nominal cell constants.
The Salt 6+ can be used with 1.0 only. Default value is k = 1.0 [C 1.0].

See Section 5.2 Advanced Setup to modify.

3.

The third screen shows the Normalization Temperature which can be set to 25 °C
or 20 °C. Default value is 25 °C [t 25.0 °C].

See Section 5.6 Advanced Setup to modify.

4.

The fourth screen shows the Temperature Coefficient which can be set from 0.0
to 3.0 % per °C. Default value is 2.1 %/°C [t 2.1%].

See Section 5.5 Advanced Setup to modify.

5.

All LCD segments light for 2 seconds before entering measurement mode. Note:
the meter will use the measurement mode that was in use when it was previously
powered off.

Cond 6+

Measurement Mode

m

S

µ

C

°

%

pptmV

pH %

MIN

MAX

LO

HO

F

C

° °

m

S

µ

S

µ

ON

OFF

This manual is related to the following products: