F4 –INITIAL ZERO RANGE

To enter this parameter, press the [Tare] key when “F4 int” is shown.

The display will show the current initial zero range.

Press the [Func] key to change the value and press [Tare] to accept the value.

Press [Zero] to return to weighing.

F5 -RE-ZERO RANGE

To enter this parameter, press the [Tare] key when “F5 rEZ” is shown.

The display will show the current re-zero range.

Press the [Func] key to change the value.

Press [Tare] to accept the value.

Press [Zero] to return to weighing.

F6 -SUCCESSIVE TARE

To enter this parameter, press the [Tare] key when “F6 SCS” is shown.

The display will show if the successive tare is on or off.

Press the [Func] key to change the value.

Press [Tare] to accept the displayed value.

Press [Zero] to return to weighing.

F7 –ADC COUNTS

To enter this parameter, press the [Tare] key when “F7 Cnt” is shown.

This parameter allows you to view the A/D counts from the internal A/D converter.
This can be an aid to service.

Press the [Tare] key to return to the PARAMETER menu.

Press the [Zero] key to return to weighing.

Typical value at zero is 30,000-90,000 (approx.)

Typical value at full capacity is 500,000 (approx.)