7.0  TROUBLE

 

SHOOTING 

7.1

 

ERROR

 

MESSAGES 

If an error message is shown, repeat the step that caused the message. If the error 
message is still shown then contact your dealer for support. 

ERROR 
CODE 

DESCRIPTION 

POSSIBLE CAUSES 
 

Solutions 

ADO

A/D Value too high. 

Weight on the pan when turning 
on.  
Improper calibration of the 
balance. 
Damaged load cell. 
Damaged Electronics. 

Remove any weight from the 
stainless steel pan.  
Recalibrate. 

ADL

A/D Value too low. 

Pan is not installed when turning 
on. 
Calibration not correct. 
Damaged load cell. 

Install pan and power back 
on.  
 
Recalibrate. 

-----

A/D count is not correct 
when turning the 
balance on. 

Load cell damaged. 
Electronics damaged. 

Remove any weight from the 
stainless steel pan and try 
rebooting. Try calibrating the 
balance.  

Unit-E

Unit error. 

  

 

Low battery indicator.  Batteries may be flat. 

Charge the battery. 

FAIL

Calibration failure. 

Incorrect mass used to calibrate. 
The user calibration is not within 
5% of factory calibration.  
Possible damaged to loadcell. 

Try to recalibrate, check that 
the weight used to calibrate 
matches that of the value of 
the balance.  

Unstable – machine 
cannot get a stable 
reading. 

Possible damage to the Loadcell. Make sure the balance is on a 

flat surface and away from 
vibration.  

 

No Power when turning 
on. 

Battery may be flat. AC adapter 
may not be working.  

Change the batteries. Or use 
the AC adapter 

 

 

 

 

 

15 | 