0 measurement of solid density – Nova-Tech Density Kit for Adam PW and PGW Series User Manual
Page 6

5.0 MEASUREMENT OF SOLID DENSITY
To perform the density of solids test, the user must have a method to
immerse the solid item in a chosen liquid. The density of the liquid must
be known or determined from a look-up table.
For water and ethanol the density will be calculated based on the
temperatures entered using the numeric entry method, when asked by
the balance.
Numeric Entry Method-
To set a value when required, use the keys as given below-
- [Up] and [Down] keys to increase or decrease the flashing digit,
- [Advance] and [Back] keys to advance or move back the digit
and
- [Enter] key to accept the value
Steps:
• After installing the Density Determination Kit, tare the balance by
pressing the [Î0/TÍ] key
• Press [Mode] and then [Up] and [Down] keys to select “Density”
• Press [Enter] to enter the Density mode
• Press the [Up] and [Down] keys to select the Solid or Liquid
method when “DENSITY SOLId”” or “DENSITY LIqUId” is
displayed
• Press [Enter] to select the “DENSITY SOLId” method
• Press [Up] or [Down] to select the liquid of choice- Water, Ethanol
or Other
• Press [Enter] to select the choice. For water and Ethanol the
temperature will be asked for. Enter the temperature using the
numeric entry method
• For the “Other” choice the density will be asked for. Enter the
© Adam Equipment Company 2007
4