Meriam M1 – Series Pressure Calibrator User Manual

Page 5

background image

M1 Series – PC

Instruction Manual 9R000035-IR

E.O. 6409 JAN 2008

www.meriam.com

Page 5 of 18

Zeroing the M1

To Zero the M1, first turn off pressure sources and vent pressure ports to
atmosphere. Then simultaneously press and hold the F and U keys until
“F.CAL” is displayed. The process is complete when the display changes
from “----” to a “zero” reading.

Note: The M1 can only be zeroed if the applied pressure is within ±1% of

the full scale range. Further details are provided under Field
Recalibration.



Set Functional Mode

There are 5 possible modes: NORMAL, HOLD, MIN/MAX, AVE and TEMP
(°F or °C). NORMAL (pressure measure mode) is the default on power up,
and is active when the latter four modes are not displayed on the LCD.

Functional Modes

HOLD

Freezes the pressure measurement on the
display, however, the bar graph is still active

MIN/MAX

Displays the MIN and MAX pressures

AVE

Displays a filtered average of the pressure
measurements

TEMP

Displays the internal ambient temperature of the
M1


The NORMAL mode has an additional sub mode referred to as “REC” or
record mode, which is described later.