Sheath fluid – Luminex 100 IS User Manual Addendum Version 2.3 User Manual
Page 8

Addendum to User Manual Version 2.3
x
MAP Technology
6
PN 89-00002-00-140 Rev. A
•
For a standard plate with flat-bottomed wells – stack two of the
larger (5.08 mm diameter) alignment discs together and place
them into the selected well.
•
For a filter-bottom plate – stack three of the larger (5.08 mm
diameter) alignment discs together and place them into the
selected well.
•
For a half-volume plate with flat-bottomed wells – stack two of
the smaller (3.35 mm diameter) alignment discs together and
place them into the selected well.
•
For a round-bottom (U-bottom) plate – stack two of the smaller
(3.35 mm diameter) alignment discs together and place them into
the selected well.
•
For a plate with conical wells – place one alignment sphere into
the selected well.
Sheath Fluid
USE OF NON-LUMINEX APPROVED SHEATH FLUID
SHALL CONSTITUTE “IMPROPER USE” AND MAY VOID
WARRANTY RIGHTS PROVIDED BY LUMINEX AND/OR
ITS AUTHORIZED PARTNER.