Luminex sd system, Pc specifications, Hardware components – Luminex 100 IS 2.1 User Installation Guide User Manual
Page 8
Luminex 100 IS 2.1 User Manual Installation Guide
x
MAP Technology
8
PN 89-00002-00-105 Rev. A
•
Weight: 15 kg (33 lbs.)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
•
Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C
•
Luminex XYP instrument, input voltage range: 100-240 V~
±10%, 1.8 Amps, 47-63 Hz
•
AC inlet fuse: 3 A, 250 V~, fast acting
Luminex SD System
•
Ambient temperature: 15°C to 30°C (59° to 86°F)
•
Humidity: 20% to 80%, noncondensing
•
Altitude: designed to operate at up to 2400m (7874 feet) above
mean sea level
•
Physical dimensions: 20 cm (8 inches) W x 30 cm (11.75 inches)
D x 24.75 cm (9.75 inches) H
•
Weight: 9 kg (20 lbs)
•
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
•
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
•
Input voltage range: 100-240 V~ ±10%, 0.4 Amps, 47-63 Hz
•
AC inlet fuse: 2 Amp, 250 V~, time lag
PC Specifications
•
CE marked and UL listed
•
Power cords: Power cords specific to the country of use are
included in the system as required
•
Power: 115-230 V~, 6 Amps, 50-60 Hz
Hardware Components
The Luminex 100 IS system includes the following hardware
components:
•
Luminex 100 analyzer
•
Computer (PC), monitor, and accessories
•
Luminex XYP instrument
•
Luminex Sheath Delivery System (Luminex SD™) (optional)
•
Power cables
•
Alignment guide
•
One short sample probe
•
Luminex XYP instrument sample probe