beautypg.com

Luminex 100 IS 2.1 User Installation Guide User Manual

Page 3

background image

x

MAP Technology

Luminex 100 IS System Hardware Setup

PN 89-00002-00-105 Rev.A

3

4. Remove the red shipping pin from the XYP instrument. Leave

the silver knob in the Luminex XYP instrument. See Figure 2.

5. Place the Luminex 100 analyzer onto the Luminex XYP

instrument. See Figure 1 and Figure 3 for positioning.

6. Install the alignment guide (translucent plastic tube) by screwing

it through the opening on the Luminex analyzer into the
Luminex XYP instrument. See Figure 3. Adjust the position of
the Luminex 100 analyzer with the Luminex XYP instrument
until the alignment guide threads screw in completely.

Figure 3 Install Alignment Guide

7. Ensure that the power switches on the XYP instrument and

Luminex 100 analyzer are in the off position.

8. Attach the power cord to the power input module of the Luminex

XYP instrument. Attach the serial cable to the Luminex XYP
instrument. Do not plug the Luminex XYP instrument into the
power outlet.

9. Attach the power cord to the input module of the Luminex 100

analyzer. Attach the Luminex 100 analyzer USB cable
(PN 85-10011-00-046) to P1 on the Luminex 100 analyzer;
reference item 1 in Figure 4. Do not plug the power cord into the
power outlet.

10. Connect the Luminex 100 analyzer USB cable (PN 85-10011-00-

046) to a USB port on the PC; reference item 13 in Figure 4.
Connect the Luminex XYP instrument serial cable to the COM
port on the PC; reference item 15 in Figure 4.

Note:

The USB cable is 5 feet

long.