3 event codes, Troubleshooting – Hach-Lange 1720E Low Range Turbidimeter User Manual
Page 35

31
Troubleshooting
7.3 Event Codes
Event codes are not displayed on controller and must be downloaded from the event log to
be viewed. Troubleshooting actions are provided in
High Readings
Deionized water turbidity is greater than
0.5 NTU
Clean the instrument.
Access Calibration History for turbidity value of
ultra-filtered water.
Verify the flow is between 200–750 mL/min.
Recalibrate the instrument.
Table 6 Additional Malfunctions Not Recorded in the Event Log (continued)
Symptom
Possible Cause
Corrective Action
Table 7 Event Log List
Event
Event #
Data1
Data2
Data3
Bubble Reject Change
0
0 = OFF
1= ON
—
—
Filter Size Change
1
0 = No avg, 1 = 6 sec
2 = 30 sec, 3 = 60 sec
4 = 90 sec
—
—
Data Log Interval Change
2
0 = 30 sec, 1 = 1min
2 = 5 min, 3 = 10 min
4 = 15 min
—
—
Power On
3
1
—
—
Calibration
4
Cal Gain
Initials
—
Verification
5
Expected Value
Meas Value
Initials
Dark Reading Warning
6
Measured Value
—
—
Temperature Warning
7
Temperature Value
—
—
Voltage Warning
8
16 = +5V high
32 = +5V low
64 = V in high
128 = V in low
—
—
Lamp Warning
9
1 = lamp V high
2 = lamp V low
4 = lamp I high
8 = lamp I low
—
—
A/D Fail
10
1
—
—
Lamp Fail
11
2
—
—
Flash Fail
12
4
—
—
Output Mode Change
13
0 = Active, 1 = Hold
2 = Transfer, 4 = Normal
—
—