1 mounting the turbidimeter body, 2 installing the head assembly, 4 installing a sample line – Hach-Lange 1720E Low Range Turbidimeter User Manual
Page 17: Installation 3.3.1 mounting the turbidimeter body

13
Installation
3.3.1 Mounting the Turbidimeter Body
Locate the turbidimeter as close to the sampling point as possible. A shorter distance for 
the sample to travel results in a faster response time. 
Clean the turbidimeter body and bubble trap before installation using the instructions 
supplied in 
. Slotted mounting brackets are integral parts of the
turbidimeter body. Install customer-supplied hardware appropriate for the installation 
environment using the criteria detailed below:
•
Install in a location that is isolated from vibration.
•
Allow at least 22 cm (approximately 10 inches) clearance for removal of the head 
assembly and bubble trap cover from the top of the turbidimeter body. 
•
Leave enough room below the turbidimeter body to remove the bottom plug and to 
place a container under the drain when calibrating or cleaning.
Note: Make sure the top of the turbidimeter body is level.
•
Install two ¼-20 bolts 10-
3
/
4
inches apart (on center). Leave at least ¼-inch of the bolt
head exposed.
•
Make sure the bolts are installed level.
Slide the slotted mounting brackets of the turbidimeter body onto the bolts.
3.3.2 Installing the Head Assembly
After the turbidimeter body has been mounted, install the bubble trap cover, then place the 
head assembly on the turbidimeter body with the label facing the front. Move the head 
assembly back and forth slightly to ensure it is properly seated on the body of the 
instrument. Failure to properly seat the head will result in light leakage and 
erroneous readings.
The rear portion of the head assembly has a molded “lip” which may be used to hang the 
head assembly on the turbidimeter body edge for routine maintenance. 
3.4 Installing a Sample Line
DANGER
This turbidimeter is not designed for use in hazardous locations or with samples 
that are flammable or explosive in nature. If any sample solution other than water is 
used in this product, test the sample/product compatibility to ensure user safety 
and proper product performance.
One-fourth inch OD rigid or semi-rigid tubing is recommended for sample lines. Run them 
as directly as possible between the turbidimeter body and the sampling point to minimize 
sample flow lag time. 
Install sample line taps into larger process pipes to minimize interference from air bubbles 
or pipeline bottom sediment. A tap projecting into the center of the pipe is ideal. 
shows examples of sample tap installations.
Note: When setting the flow rate, take care to avoid sweeping air “micro-bubbles” through the 
internal bubble trap. Observe the sample flow inside the turbidimeter body. If small air bubbles can 
be seen flowing up through the center, reduce the flow rate.
